HI93703 Turbidity Meter ISO 7027 Compliant

Product Code : HI93703

Category : Food&Diary

Catalog : sorry not ready

Product information "HI93703 Turbidity Meter ISO 7027 Compliant"

GLP Features
3 point calibration



HI 93703 series are microprocessor based turbidity meters that provide laboratory accuracy for field turbidity measurements.

The HI 93703 is easy to calibrate (with 3 point calibration) using AMCO-EPA standards and is ISO 7027 compliant.

Calibration Standards are optional.



It meets the requirements of GLP (Good Laboratory Practice) protocols for last calibration data storage.

* FTU is equivalent to the other internationally recognized unit NTU (Nephelometric Turbidity Unit.

Order Information: HI 93703 is supplied complete with glass cuvette, batteries and instructions.


Specification :
Range 0.00 to 50.00 FTU; 50 to 1000 FTU
Resolution 0.01 FTU (0.00 to 50.00 FTU); 1 FTU (50 to 1000 FTU)
Accuracy ±0.5 FTU or ±5% of reading (whichever is greater)
Calibration 3 points (0 FTU, 10 FTU and 500 FTU)
Light Source infrared LED
Light Detection Silicon Photocell
Battery Type / Life 1.5V AA (4) /approximately 60 hours of continuous use or 900 measurements; auto-off after 5 minutes of non-use
Environment 0 to 50°C (32 to 122°F); RH max 95% non-condensing
Dimensions 220 x 82 x 66 mm (8.7 x 3.2 x 2.6")
Weight 510 g (1.1 lb.)