HI98196 Multiparameter - pH / ORP / DO / Temp - 4m Cable Complete Set

Product Code : HI98196

Category : Environmental

Catalog : sorry not ready

Product information "HI98196 Multiparameter - pH / ORP / DO / Temp - 4m Cable Complete Set"
Rugged, waterproof and easy to use, the HI 98196 is the ideal meters for field measurements of lakes, rivers and seas.
The HI 98196 can measure water quality parameters such as pH, ORP, temperature and dissolved oxygen concentration. Atmospheric pressure is measured for DO compensation.
HI 9819X incorporates comprehensive GLP features and the transferring of data via USB connection. The HI 9819X HELP key displays context sensitive help. The alpha-numeric keypad offers a user friendly way to complete the input fields.
Specification :
pH Range 0.00 to 14.00 pH
pH Resolution 0.01 pH
pH Accuracy ±0.02 pH
pH Calibration automatic one, two, or three points with automatic recognition of five standard buffers (pH 4.01, 6.86, 7.01, 9.18, 10.01) or one custom buffer
mV Range ±600.0 mV
mV Resolution 0.1 mV
mV Accuracy ±0.5 mV
ORP Range ±2000 mV
ORP Resolution 0.1 mV
ORP Accuracy ±1.0 mV
ORP Calibration automatic at one custom point (relative mV)
DO Range 0.0 to 500.0%; 0.00 to 50.00 ppm (mg/L)
DO Resolution 0.1%; 0.01 ppm (mg/L)
DO Accuracy 0.0 to 300.0% saturation: ±1.5% of reading or ±1.0% saturation whichever is greater, 0.00 to 30.00 ppm (mg/L): ±1.5% of reading or ±0.10 ppm (mg/L), whichever is greater, 30.00 ppm (mg/L) to 50.00 ppm (mg/L): ±3% of reading, 300.0 to 500.0% saturation: ±3% of reading
DO Calibration automatic one or two point at 100% saturation (8.26 mg/L) and 0% saturation (0 mg/L).
Atmospheric Pressure Range 450 to 850 mm Hg; 17.72 to 33.46 in Hg; 600.0 to 1133.2 mbar; 8.702 to 16.436 psi; 0.5921 to 1.1184 atm; 60.00 to 113.32 kPa
Atmospheric Pressure Resolution 0.1 mm Hg; 0.01 in Hg; 0.1 mbar; 0.001 psi; 0.0001 atm; 0.01 kPa
Atmospheric Pressure Accuracy ±3 mm Hg within ±15°C from the temperature during calibration
Atmospheric Pressure Calibration automatic at one custom point
Temperature Range -5.00 to 55.00 °C, 23.00 to 131.00 °F, 268.15 to 328.15 K
Temperature Resolution 0.01 K, 0.01 °C, 0.01 °F
Temperature Accuracy ±0.15 °C; ±0.27 °F; ±0.15 K
Temperature Calibration automatic at one custom point
Temperature Compensation automatic from -5 to 55 °C (23 to 131 °F)
Logging Memory 45,000 records (continuous logging or log-on-demand of all parameters)
Logging Interval one second to three hours
PC Connectivity USB to PC with HI9298194 Hanna software installed
Battery Type/Life 1.5V AA batteries (4) / approximately 360 hours of continuous use without backlight (50 hours with backlight) With probe connected, approximately 104 hours of continuous use without backlight.
Environment 0 to 50°C (32 to 122°F); RH 100% (IP67)
Dimensions 185 x 93 x 35.2 mm (7.3 x 3.6 x 1.4”)
Weight 400 g (14.2 oz.)
Ordering Information HI98196 is supplied with HI7698196 probe, HI7698194-1 pH/ORP sensor, HI7698194-2 DO sensor, HI 76981942 Probe maintenance kit (HI 7042S DO electrolyte solution, DO membrane caps (5), O-rings for sensors (5), syringe with grease to lubricate the O-rings), HI9298194 Hanna PC software, HI920015 micro USB cable, 1.5V AA batteries (4), instruction manual, quick start guide, quality certificate, and rugged carrying case with custom