HI9835 EC / TDS / NaCl / °C-Meter, Auto-Endpoint, ATC

Product Code : HI9835

Category : Plating

Catalog : sorry not ready

Product information "HI9835 EC / TDS / NaCl / °C-Meter, Auto-Endpoint, ATC"

HI 9835 is handheld conductivity/TDS/salinity/temperature meters. Users are provided with a series of diagnostic features and messages on the LCD which guide the user through calibration, operation and troubleshooting.

Conductivity and TDS measurement parameters are selectable such as: cell constant range from 0.500 to 1.700, temperature coefficient from 0.00 to 6.00%/ºC, temperature reference from 20 to 25ºC and a selectable TDS factor of 0.40 to 0.80.

This instrument utilizes the four ring HI 76309/1.5 conductivity probe with internal temperature sensor. The four ring design offers accurate readings over all the conductivity range with immunity to polarization and falling that occurs on long term use of amperometric probes.

The autoranging feature of the EC and TDS modes automatically sets the meter to the scale with the highest possible resolution. The auto endpoint feature automatically freezes the display once a stable reading is reached.



Autoranging, manual ranging and range lock
Four ring probe design
Automatic, manual and no Temperature Compensation
Auto endpoint; Automatically freezes stable readings on the LCD display
GLP features
BEPS; (Battery Error Prevention System) alerts the user in the event that low battery power could adversely affect readings
On-screen user guides
Backlit display
% Battery displayed on startup



Order Information: HI 9835 is supplied with HI 76309 conductivity probe, batteries, instructions and rugged carrying case.


Specification :
Range EC 0.00 to 29.99 µS/cm; 30.0 to 299.9 µS/cm; 300 to 2999 µS/cm; 3.00 to 29.99 mS/cm; 30.0 to 200.0 mS/cm; up to 500.0 mS/cm (actual EC)*
TDS 0.00 to 14.99 mg/L (ppm); 15.0 to 149.9 mg/L (ppm); 150 to 1499 mg/L (ppm); 1.50 to 14.99 g/L (ppt); 15.0 to 100.0 g/L (ppt); up to 400.0 g/L (ppt) (actual TDS)* with 0.80 conversion factor
NaCl 0.0 to 400.0%
Temperature –20.0 to 120.0 ºC (-4.0 to 248.0ºF)
Resolution EC 0.01 µS/cm; 0.1 µS/cm; 1 µS/cm; 0.01 mS/cm; 0.1 mS/cm
TDS 0.01 mg/L (ppm); 0.1 mg/L (ppm); 1 mg/L (ppm); 0.01 g/L (ppt); 0.1 g/L (ppt)
NaCl 0.1%
Temperature 0.1°C
Accuracy @ 25°C EC ±1 % of reading (±0.05 µS/cm or 1 digit)
TDS ±1 % of reading (±0.03 mg/L (ppm) or 1 digit, whichever greater)
NaCl ±1% of reading
Temperature ±0.2°C (excluding probe error)
Calibration EC automatic, one point with six memorized values (84, 1413, 5000, 12880, 80000, 111800 µS/cm)
NaCl one point with HI 7037 calibration solution
Temperature two point, at 0 and 50°C (32 and 122°F)
Temperature Compensation automatic or manual from –20.0 to 120.0 ºC (-4.0 to 248.0ºF) (can be disabled for measuring conductivity)
Temperature Coefficient selectable from 0.00 to 6.00%/ºC (EC and TDS only); default value is 1.90%/ºC
TDS Factor selectable from 0.40 to 0.80 (default value is 0.50)
Probe HI 76309 four ring conductivity probe with internal temperature sensor, DIN connector and 1 m (3.3’) cable (included)
Power Supply 1.5V AAA batteries (3) /approximately 200 hours of continuous use without backlight (50 hours with backlight on); auto-off after 5, 10, 20 and 60 minutes (can be disabled)
Environment 0 to 50°C (32 to 122°F); RH max 95%
Dimensions 185 x 72 x 36 mm (7.3 x 2.8 x 1.4”)
Weight 300 g (10.6 oz)