HI83326 Pool and Spa Photometer and pH Meter

Product Code : HI83326

Category : Bromine

Catalog : sorry not ready

Product information "HI83326 Pool and Spa Photometer and pH Meter"
Details

The HI83326 benchtop photometer measures 11 different key water quality parameters using 12 different methods. This photometer features an innovative optical system that uses LEDs, narrow band interference filters, focusing lens and both a silicon photodetector for absorbance measurement and a reference detector to maintain a consistent light source ensures accurate and repeatable photometric readings every time.

Made with the pool and spa industry in mind, a basic necessity of pool water treatment is to maintain the water in a safe and pleasant condition for the swimmers. In pool and spa water treatment, disinfection is essential to rid the pool of bacteria and control nuisance organisms like algae which may occur in the pool, spa, filtration equipment, or piping. There are a number of available disinfectant compounds including chlorine, bromine, and ozone. In order to achieve ideal water conditions, water requires testing on a daily and sometimes hourly basis to ensure there is enough residual disinfectant and to maintain pH levels. Equally important is calcium hardness and alkalinity; these levels should be monitored weekly to ensure the pool or spa water is well balanced to avoid corrosion and scale formation.

A digital pH electrode input allows the user to measure pH by a traditional glass electrode. The digital pH electrode has a built in microchip within the probe that stores all of the calibration information. Having the calibration information stored in the probe allows for hot swapping of pH electrodes without having to recalibrate. All pH measurements are automatically compensated for temperature variations with a built in thermistor located in the tip of the sensing bulb for fast and accurate temperature measurement.

The HI83326 offers an absorbance measuring mode that allows for CAL Check standards to be used to validate the performance of the system. The absorbance mode allows the user to select one of the three wavelengths of light (525 nm, 575 nm, and 610 nm) to measure and plot their own concentration versus absorbance mode. This is useful for users with their own chemical method and for educators to teach the concept of absorbance by using the Beer-Lambert Law.

Two USB ports are provided for transferring data to a flash drive or computer and to use as a power source for the meter. For added convenience and portability the meter can also operate on an internal 3.7 VDC Lithium-polymer rechargeable battery.
pH
pH Range Photometer: 6.5 to 8.5 pH
pH electrode: -2.00 to 16.00 pH
pH Resolution Photometer: 0.1 pH
pH electrode: 0.01 pH
pH Accuracy Photometer: ±0.1 pH
pH electrode: ±0.01 pH
pH Calibration Automatic one or two point calibration with one set of standard buffers available (4.01, 6.86, 7.01, 9.18, 10.01)
pH Temperature Compensation Automatic (-5.0 to 100.0 oC; 23.0 to 212.0 oF); limits reduced based on the pH electrode used
pH CAL Check (electrode diagnostics) clean electrode and check buffer/check probe displayed during calibration
pH Method Photometer: phenol red
pH-mV Range ±1000 mV
pH-mV Resolution 0.1 mV
pH-mV Accuracy ±0.2 mV


Absorbance
Absorbance Range 0.000 to 4.000 Abs
Absorbance Resolution 0.001 Abs
Absorbance Accuracy +/-0.003Abs @ 1.000 Abs


Alkalinity
Alkalinity Range Freshwater: 0 to 500 mg/L (as CaCO3)
Alkalinity Resolution 1 mg/L
Alkalinity Accuracy ±5 mg/L ±5% of reading
Alkalinity Method Colorimetric method


Bromine
Bromine Range 0.00 to 8.00 mg/L (as Br2)
Bromine Resolution 0.01 mg/L
Bromine Accuracy ±0.08 mg/L ±3% of reading
Bromine Method Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, DPD method.


Chlorine
Free Chlorine Range 0.00 to 5.00 mg/L (as Cl2)
Free Chlorine Resolution 0.01 mg/L
Free Chlorine Accuracy ±0.03 mg/L ±3% of reading
Total Chlorine Range 0.00 to 5.00 mg/L (as Cl2)
Total Chlorine Resolution 0.01 mg/L
Total Chlorine Accuracy ±0.03 mg/L ±3% of reading
Chlorine Method Adaptation of the EPA 330.5 DPD method


Copper
Copper Range High range: 0.00 to 5.00 mg/L (as Cu2+)
Copper Resolution 0.01 mg/L
Copper Accuracy High Range ±0.02 mg/L ±4% of reading
Copper Method Adaptation of the EPA bicinchoninate method


Cyanuric Acid
Cyanuric Acid Range 0 to 80 mg/L (as CYA)
Cyanuric Acid Resolution 1 mg/L
Cyanuric Acid Accuracy ±1 mg/L ±15% of reading
Cyanuric Acid Method Adaptation of the turbidimetric method


Hardness, Calcium
Hardness, Calcium Range 0.00 to 2.70 mg/L (as CaCO3)
Hardness, Calcium Resolution 0.01 mg/L
Hardness, Calcium Accuracy ±0.11 mg/L ±5% of reading
Hardness, Calcium Method Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Calmagite method


Iron
Iron Range High Range: 0.00 to 5.00 mg/L (as Fe)
Iron Resolution 0.01 mg/L
Iron Accuracy High Range: ±0.04 mg/L ±2% of reading
Iron Method High Range: Adaptation of the EPA Phenanthroline method 315B, for natural and treated waters


Nitrate
Nitrate Range 0.0 to 30.0 mg/L (as NO3-- N)
Nitrate Resolution 0.1 mg/L
Nitrate Accuracy ±0.5 mg/L ±10% of reading
Nitrate Method Adaptation of the cadmium reduction method


Ozone
Ozone Range 0.00 to 2.00 mg/L (as O3)
Ozone Resolution 0.01 mg/L
Ozone Accuracy ±0.02 mg/L ±3% of reading
Ozone Method Colorimetric DPD Method


Phosphate
Phosphate Range Freshwater
Low Range: 0.00 to 2.50 mg/L (as PO43-)
Phosphate Resolution Freshwater: 0.01 mg/L
Phosphate Accuracy Freshwater
Low Range: ±0.04 mg/L ±4% of reading
Phosphate Method Freshwater Low Range: Adaptation of the Ascorbic Acid method


General Meter
Input Channels 1 pH electrode input and 3 photometer wavelengths
pH Electrode digital pH electrode (not included)
Logging Type log on demand with user name and sample ID optional input
Logging Memory 1000 readings
Connectivity USB-A host for flash drive; micro-USB-B for power and computer connectivity
GLP calibration data for connected pH electrode
Display 128 x 64 pixel LCD with backlight
Battery Type / Life 3.7 VDC Li-polymer rechargeable battery / >500 photometric measurements or 50 hours of continuous pH measurement
Power Supply 5 VDC USB 2.0 power adapter with USB-A to micro-USB-B cable (included)
Environment 0 to 50.0 oC (32 to 122.0 oF); 0 to 95% RH, non-condensing
Dimensions 206 x 177 x 97 mm (8.1 x 7.0 x 3.8")
Weight 1.0 kg (2.2 lbs.)
Photometer/Colorimeter Light Source 3 LEDs with 525 nm, 575 nm, and 610 nm narrow band interference filters
Photometer/Colorimeter Light Detector silicon photodetector
Bandpass Filter Bandwidth 8 nm
Bandpass Filter Wavelength Accuracy ±1 nm
Cuvette Type round, 24.6 mm
Number of Methods 128 max.
Ordering Information HI83226 is supplied with sample cuvettes and caps (4 ea.), cloth for wiping cuvettes, USB to micro USB cable connector, power adapter and instruction manual.