HI83306 Environmental Analysis Photometer and pH Meter

Product Code : HI83306

Category : pH (reagents)

Catalog : sorry not ready

Product information "HI83306 Environmental Analysis Photometer and pH Meter"
Details

The HI83306 benchtop photometer measures 16 different key water quality parameters using 23 different methods. This photometer features an innovative optical system that uses LEDs, narrow band interference filters, focusing lens and both a silicon photodetector for absorbance measurement and a reference detector to maintain a consistent light source ensures accurate and repeatable photometric readings every time.

The HI83306 was developed to measure the most common parameters in environmental water quality monitoring. Nutrients such as nitrates and phosphates are key indicators of nutrient pollution from agricultural sources and are considered dangerous to environmental waters.Too few nutrients and waters will be unable to sustain healthy ecosystems; too many nutrients and algal blooms can form, which can be detrimental to water quality and aquatic health. Dissolved oxygen is an essential to performing biological processes for many forms of aquatic life, such as fish, plants and microorganisms.

A digital pH electrode input allows the user to measure pH by a traditional glass electrode. The digital pH electrode has a built in microchip within the probe that stores all of the calibration information. Having the calibration information stored in the probe allows for hot swapping of pH electrodes without having to recalibrate. All pH measurements are automatically compensated for temperature variations with a built in thermistor located in the tip of the sensing bulb for fast and accurate temperature measurement.

The HI83306 offers an absorbance measuring mode that allows for CAL Check standards to be used to validate the performance of the system. The absorbance mode allows the user to select one of the four wavelengths of light (420 nm, 525 nm, 575 nm, and 610 nm) to measure and plot their own concentration versus absorbance mode. This is useful for users with their own chemical method and for educators to teach the concept of absorbance by using the Beer-Lambert Law.

Two USB ports are provided for transferring data to a flash drive or computer and to use as a power source for the meter. For added convenience and portability the meter can also operate on an internal 3.7 VDC Lithium-polymer rechargeable battery.
pH
pH Range Photometer: 6.5 to 8.5 pH
pH electrode: -2.00 to 16.00 pH
pH Resolution Photometer: 0.1 pH
pH electrode: 0.01 pH
pH Accuracy Photometer: ±0.1 pH
pH electrode: ±0.01 pH
pH Calibration Automatic one or two point calibration with one set of standard buffers available (4.01, 6.86, 7.01, 9.18, 10.01)
pH Temperature Compensation Automatic (-5.0 to 100.0 oC; 23.0 to 212.0 oF); limits reduced based on the pH electrode used
pH CAL Check (electrode diagnostics) clean electrode and check buffer/check probe displayed during calibration
pH Method Photometer: phenol red
pH-mV Range ±1000 mV
pH-mV Resolution 0.1 mV
pH-mV Accuracy ±0.2 mV


Oxygen, Dissolved
Oxygen, Dissolved Range 0.0 to 10.0 mg/L (as O2)
Oxygen, Dissolved Resolution 0.1 mg/L
Oxygen, Dissolved Accuracy ±0.4 mg/L ±3% of reading
Oxygen, Dissolved Method Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Azide modified Winkler method


Absorbance
Absorbance Range 0.000 to 4.000 Abs
Absorbance Resolution 0.001 Abs
Absorbance Accuracy +/-0.003Abs @ 1.000 Abs


Ammonia
Ammonia Range Low Range: 0.00 to 3.00 mg/L
Medium Range: 0.00 to 10.00 mg/L
High Range: 0.0 to 100.0 mg/L (all as NH3-N)
Ammonia Resolution Low and Medium Range: 0.01 mg/L
High Range: 0.1 mg/L
Ammonia Accuracy Low Range: ±0.04 mg/L ±4% of reading
Medium Range: ±0.05 mg/L ±5% of reading
High range: ±0.5 mg/L ±5% of reading
Ammonia Method Adaptation of the ASTM Manual of Water and Environmental Technology, D1426-92, Nessler method


Chlorine
Free Chlorine Range 0.00 to 5.00 mg/L (as Cl2)
Free Chlorine Resolution 0.01 mg/L
Free Chlorine Accuracy ±0.03 mg/L ±3% of reading
Total Chlorine Range 0.00 to 5.00 mg/L (as Cl2)
Total Chlorine Resolution 0.01 mg/L
Total Chlorine Accuracy ±0.03 mg/L ±3% of reading
Chlorine Method Adaptation of the EPA 330.5 DPD method


Chromium, Hexavalent
Chromium, Hexavalent Range Low Range: 0 to 300 ?g/L (as Cr6+)
High Range: 0 to 1000 ?g/L (as Cr6+)
Chromium, Hexavalent Resolution 1 ?g/L
Chromium, Hexavalent Accuracy Low Range: ±1 ?g/L ±4% of reading
High Range: ±5 ?g/L ±4% of reading
Chromium, Hexavalent Method Adaptation of the ASTM Manual of Water and Environmental Technology, D1687-92, Diphenylcarbohydrazide method.


Color, Water
Color, Water Range 0 to 500 PCU (Platinum Cobalt Units)
Color, Water Resolution 1 PCU
Color, Water Accuracy ±10 PCU ±5% of reading
Color, Water Method Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Colorimetric Platinum Cobalt method.


Copper
Copper Range Low Range: 0.000 to 1.500 mg/L (as Cu2+)
High range: 0.00 to 5.00 mg/L (as Cu2+)
Copper Resolution 0.001 mg/L; 0.01 mg/L
Copper Accuracy Low Range: ±0.01 mg/L ±5% of reading
High Range ±0.02 mg/L ±4% of reading
Copper Method Adaptation of the EPA bicinchoninate method


Cyanuric Acid
Cyanuric Acid Range 0 to 80 mg/L (as CYA)
Cyanuric Acid Resolution 1 mg/L
Cyanuric Acid Accuracy ±1 mg/L ±15% of reading
Cyanuric Acid Method Adaptation of the turbidimetric method


Molybdenum
Molybdenum Range 0.0 to 40.0 mg/L (as Mo6+)
Molybdenum Resolution 0.1 mg/L
Molybdenum Accuracy ±0.3 mg/L ±5% of reading
Molybdenum Method Adaptation of the mercaptoacetic acid method


Nickel
Nickel Range Low Range: 0.000 to 1.000 mg/L (as Ni)
High Range: 0.00 to 7.00 g/L (as Ni)
Nickel Resolution 0.001 mg/L; 0.01 g/L
Nickel Accuracy Low range: ±0.010 mg/L ±7% of reading
High Range: ±0.07 g/L ±4% of reading
Nickel Method Low Range: Adaptation of the PAN method
High Range: Adaptation of the photometric method


Nitrate
Nitrate Range 0.0 to 30.0 mg/L (as NO3-- N)
Nitrate Resolution 0.1 mg/L
Nitrate Accuracy ±0.5 mg/L ±10% of reading
Nitrate Method Adaptation of the cadmium reduction method


Nitrite
Nitrite Range 0 to 150 mg/L (as NO2-)
Nitrite Resolution 1 mg/L
Nitrite Accuracy ±4 mg/L ±4% of reading
Nitrite Method Adaptation of the Ferrous Sulfate method


Phosphate
Phosphate Range Freshwater
Low Range: 0.00 to 2.50 mg/L (as PO43-)
High range: 0.0 to 30.0 mg/L(as PO43-)
Phosphate Resolution 0.01 mg/L; 0.1 mg/L
Phosphate Accuracy Low Range: ±0.04 mg/L ±4% of reading
High Range: ±1 mg/L ±4% of reading
Phosphate Method Low Range: Adaptation of the Ascorbic Acid method
High Range: Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Amino Acid method


Silica
Silica Range 0.00 to 2.00 mg/L (as SiO2)
Silica Resolution 0.01 mg/L
Silica Accuracy ±0.03 mg/L ±3% of reading
Silica Method Adaptation of the ASTM Manual of Water and Environmental Technology, D859, Heteropoly Molybdenum Blue method


Silver
Silver Range 0.000 to 1.000 mg/L (as Ag)
Silver Resolution 0.001 mg/L
Silver Accuracy ±0.020 mg/L ±5% of reading
Silver Method Adaptation of the PAN method


Zinc
Zinc Range 0.00 to 3.00 mg/L (as Zn)
Zinc Resolution 0.01 mg/L
Zinc Accuracy ±0.03 mg/L ±3% of reading
Zinc Method Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Zincon method


General Meter
Input Channels 1 pH electrode input and 4 photometer wavelengths
pH Electrode digital pH electrode (not included)
Logging Type log on demand with user name and sample ID optional input
Logging Memory 1000 readings
Connectivity USB-A host for flash drive; micro-USB-B for power and computer connectivity
GLP calibration data for connected pH electrode
Display 128 x 64 pixel LCD with backlight
Battery Type / Life 3.7 VDC Li-polymer rechargeable battery / >500 photometric measurements or 50 hours of continuous pH measurement
Power Supply 5 VDC USB 2.0 power adapter with USB-A to micro-USB-B cable (included)
Environment 0 to 50.0 oC (32 to 122.0 oF); 0 to 95% RH, non-condensing
Dimensions 206 x 177 x 97 mm (8.1 x 7.0 x 3.8")
Weight 1.0 kg (2.2 lbs.)
Photometer/Colorimeter Light Source 4 LEDs with 420 nm, 525 nm, 575 nm, and 610 nm narrow band interference filters
Photometer/Colorimeter Light Detector silicon photodetector
Bandpass Filter Bandwidth 8 nm
Bandpass Filter Wavelength Accuracy ±1 nm
Cuvette Type round, 24.6 mm
Number of Methods 128 max.
Ordering Information HI83306 is supplied with sample cuvettes and caps (4 ea.), cloth for wiping cuvettes, USB to micro USB cable connector, power adapter and instruction manual.