HI6222-02 pH/ORP/ISE Benchtop Meter Starter Kit
Product Code : HI6222-02
Category : Benchtop
Catalog : hi6222-02_ph_orp_ise_benchtop_meter_starter_kit_support.pdf
HI6222-02 Multiparameter Modular pH/ORP/ISE Benchtop Meter
Highlight Utama
- Mengukur pH, ORP, Ion Selective Electrode (ISE), mV, dan suhu.
- Dilengkapi dua modul HI6000-2 pH/ORP/ISE yang telah terpasang.
- Mendukung dua kanal pengukuran pH/ORP/ISE dalam satu instrumen.
- Menggunakan sistem modular HI6000 yang dapat dikembangkan sesuai kebutuhan laboratorium.
- Layar sentuh kapasitif 7 inci dengan tampilan hingga tiga kanal secara bersamaan.
- Mendukung metode ISE Direct, Known Addition, Known Subtraction, Analyte Addition, dan Analyte Subtraction.
- Penyimpanan hingga 1.000.000 data dengan konektivitas USB, Ethernet, Wi-Fi, FTP, dan Email.
Deskripsi Produk
HI6222-02 Multiparameter Modular pH/ORP/ISE Benchtop Meter merupakan sistem pengukuran laboratorium berbasis platform modular HI6000 yang dirancang untuk analisis pH, ORP, Ion Selective Electrode (ISE), mV, dan suhu. Paket ini dilengkapi dengan dua modul HI6000-2 pH/ORP/ISE yang telah terpasang, sehingga pengguna dapat melakukan pengukuran melalui dua kanal secara fleksibel dalam satu instrumen.
Dengan konfigurasi dua modul pH/ORP/ISE, HI6222-02 sangat sesuai untuk laboratorium yang perlu membandingkan dua sampel, menggunakan dua jenis elektroda, atau menjalankan pengukuran pH dan ISE tanpa harus mengganti koneksi sensor secara berulang. Setiap modul mendukung pengukuran pH, mV, relatif mV, serta berbagai satuan konsentrasi ISE seperti ppm, ppb, mg/L, g/L, mol/L, mmol/L, %w/v, dan satuan yang ditentukan pengguna.
Untuk analisis ion, HI6222-02 mendukung pengukuran langsung serta metode inkremental seperti Known Addition, Known Subtraction, Analyte Addition, dan Analyte Subtraction. Metode ini berguna untuk sampel yang kompleks atau memiliki kekuatan ionik tinggi, termasuk sampel lingkungan, pertanian, farmasi, bioteknologi, makanan, air minum, dan air limbah. Hasil setiap analisis dapat disimpan sebagai laporan dan diunduh untuk dokumentasi.
Instrumen dilengkapi layar sentuh kapasitif berwarna 7 inci yang dapat menampilkan hasil pengukuran, informasi kalibrasi, status elektroda, grafik, dan tabel data. Sistem logging mampu menyimpan hingga 1.000.000 data per pengguna, sedangkan data dapat diekspor melalui USB, Ethernet, Wi-Fi, FTP, web server, atau email dalam format CSV.
Industri yang Menggunakan HI6222-02
- Industri farmasi
- Industri makanan dan minuman
- Industri kimia
- Laboratorium lingkungan
- Pengolahan air minum
- Pengolahan air limbah
- Industri pertanian dan pupuk
- Bioteknologi
- Universitas dan lembaga penelitian
- Laboratorium Quality Control
Aplikasi Umum
- Pengukuran pH pada dua sampel atau dua kanal
- Analisis ORP atau Redox
- Analisis ion menggunakan Ion Selective Electrode
- Pengukuran fluoride, chloride, nitrate, ammonia, sodium, calcium, dan ion lainnya
- Analisis sampel dengan kekuatan ionik tinggi
- Quality Control bahan baku dan produk
- Analisis air minum dan air limbah
- Penelitian dan pengembangan metode
- Pengujian lingkungan dan pertanian
Kelengkapan Produk
HI6222-02 dilengkapi dengan:
- Unit utama HI6000 Multiparameter Modular System
- Dua modul HI6000-2 pH/ORP/ISE yang telah terpasang
- Elektroda pH HI1131B
- Probe suhu HI7662-TW
- Buffer pH 4,01 sebanyak 2 sachet
- Buffer pH 7,01 sebanyak 4 sachet
- Buffer pH 10,01 sebanyak 2 sachet
- Larutan pembersih elektroda sebanyak 2 sachet
- Larutan penyimpanan elektroda pH/ORP 25 mL
- Larutan elektrolit KCl 3,5 M sebanyak 30 mL
- Pipet kapiler
- Holder elektroda dengan base plate, screw, dan cable holder clip
- Adaptor daya 24 VDC
- Kabel USB-C ke USB-A
- Quick Reference Guide instrumen
- Quick Reference Guide untuk modul
- Sertifikat kualitas modul, probe, dan instrumen
| Specification Name | Detail |
|---|---|
| Sku | HI6222-02 |
| pH Range | −2.0 to 20.0 pH; −2.00 to 20.00 pH; −2.000 to 20.000 pH |
| pH Resolution | 0.1 pH; 0.01 pH; 0.001 pH |
| pH Accuracy | ±0.1 pH; ±0.01 pH; ±0.002 pH (±1 last significant digit) |
| Calibration Points | Up to 5 |
| Calibration Type | Automatic; Semiautomatic; Manual |
| Standard Buffers | Hanna and NIST (pH 1.68, 3.00, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45) |
| Custom Buffers | Up to 5 |
| Custom Group | Option to select from eight standard buffers and user-defined custom buffers |
| First Calibration Point | Offset or Point (user setting) |
| Isopotential Point | −2.000 to 20.000 pH |
| mV Range | −2000.0 to 2000.0 mV |
| mV Resolution | 1 mV; 0.1 mV |
| mV Accuracy | ±0.2 mV ±1 last significant digit |
| mV Calibration | Single point offset, ±2000.0 mV |
| ISE Range | 1.0×10¯5 to 300.0 ppt (g/L or mg/mL) 5.0×10¯3 to 1.0×105 ppm (mg/L or µg/mL) 1.0 to 5.0×107 ppb (µg/L) 1.0×10¯7 to 10.0 M (mol/L) 1.0×10¯4 to 1.0×104 mmol/L 1.0×10¯6 to 60.0 %w/v 5.0×10¯7 to 5.0×107 user |
| ISE Resolution | 1, 2, 3 significant digits |
| ISE Accuracy | ±0.5% (monovalent ions); ±1% (divalent ions) |
| ISE Calibration Points | Up to 5 |
| ISE Calibration Type | All Standards; Standard Group |
| ISE Standards | 7 standard solutions available for each concentration unit |
| ISE Custom Standards | Up to 5 |
| ISE Custom Group | Up to 5 |
| Temperature Range | −20.0 to 120.0 °C; −4.0 to 248.0 °F; 253.2 to 393.2 K |
| Temperature Resolution | 0.1 °C; 0.1 °F; 0.1 K |
| Temperature Accuracy | ±0.2 °C; ±0.4 °F; ±0.2 K |
| Temperature Calibration | Single point, adjustable |
| Temperature Compensation | Automatic; Manual |
| Reading Mode | Direct Direct/Autohold ISE Only: Known Addition, Known Subtraction, Analyte Addition, Analyte Subtraction |
| Basic View | Measurement data; Measurement profile (if enabled); Stability status |
| Simple GLP View | Basic view information; Last calibration date, slope, offset (pH, Rel. mV – ISE only) |
| Full GLP View | Simple GLP information; Calibration point details (pH & ISE) |
| Graph View | Basic view information; Measurement versus time graph |
| Table View | Basic view information, Table with measurements updated every second |
| Reading Stability Criteria | Accurate; Medium; Fast |
| Reading Sampling Rate | 1000 ms |
| Logging Type | Automatic; Manual; Autohold |
| Logging Number of Records | 50,000 maximum per file and stores at least 1,000,000 data points per user |
| Logging Automatic Interval | 1, 2, 5, 10, 30 seconds and 1, 2, 5, 10, 15, 30, 60, 120, 150, 180 minutes |
| Logging Sample ID | Incremental Mode; Manual |
| Logging Export Option | .CSV file format |
| Connectivity USB-A | 2 ports keyboard and/or printer input USB Flash Drive |
| Connectivity USB-C | 1 port PC Connectivity and USB-C Type Flash Drive |
| Connectivity Wi-Fi & Ethernet | FTP; Web Server (Log transfer and download); Email |
| Connectivity RS232 | Connecting peripherals |
| Calibration Reminder | Daily (0 minutes to 23 hours and 59 minutes); Periodic (1 minute to 30 Days, 23 hours and 59 minutes); or Disabled |
| Users | Up to 9 users and the default administrator account |
| Power Supply | DC adapter 100-240 VAC to 24 VDC 2A |
| Environment | 0 to 50 °C / 32 to 122 °F / 273 to 323 K maximum 95 % RH non-condensing |
| Dimensions | 205 x 160 x 77 mm (8.0 x 6.2 x 3.0 “) |
| Weight | 8.05 lbs |